No. 99MAH040B
・ Repeatability in the range of 0.2 mm (.0079”) from the bottom of the stroke is not
guaranteed for this indicator. When setting the origin, be sure to lift the spindle at least
0.2 mm (.0079”) from the bottom of the stroke.
・ The preset value will be calculated automatically according to the resolution. Check the
preset value when the resolution is changed since the conversion error may occur.
・ Press the upper left key to cancel the setting,
・ The preset setting cannot be completed while the spindle is moving. So complete it
after the spindle stops.
・ The power supply cannot be turned off with the upper left key in the preset setting.
・ When auto-off is activated during preset setting, the setting will be canceled.
・ The setup preset values and position will be held after the power is turned off.
However, when the battery is replaced, the preset position will be cleared. Set up the
position again.
・ The preset setting cannot be completed if Overflow error of preset value (Err95) is
occurring for the selected preset number. Reset the preset value.
・ When preset setting is changed, all the measurement values saved with the data save
function will be deleted.
・ A rubber damper has been attached to the spindle in this indicator as a shock
absorber. The elasticity of the damper may cause the indicated value to not stabilize at
the bottom of the stroke, but this will not cause any operational problems.
Also, the spindle may feel heavy at the bottom of the stroke when this indicator is first
used, but this can be resolved by pushing the spindle up once.