No. 99MAH040B
5-1
5 ERROR MESSAGE AND
ACTION
This chapter describes the error messages and their corrective
actions.
If any error occurs in this Digimatic indicator, a corresponding error message is displayed.
If the instrument does not recover the normal conditions after corrective actions, contact
your dealer or Mitutoyo sales office.
Display
definition
Actions
Low battery alarm
Battery voltage is getting
low
Replace the battery with the
new one.
Low battery error
Measurement cannot be
performed due to
consumption of the
battery.
Replace the battery with the
new one.
Sensor signal synthesizing
error
Synthesizing of sensor
signal is failed.
A sensor signal
synthesizing error occurred
while the spindle was
moving at high speed.
Keep on using the indicator
since this error does not
affect measured values.
* If this error occurs while the
spindle is stopped, it could
be due to sensor failure. In
this case, contact the
nearest Mitutoyo sales
office.