No. 99MAH040B
Overflow error of display
value
・ The display value has
exceeded the maximum
number of digits that can
be displayed.
・ Return the display value to
the permissible number of
digits to automatically reset
the error.
・ Press the center key, or
press and hold
・ Move the Setup mode to
change the resolution.
(Refer to "3.3.2 RES:
Resolution".)
Internal connection error
・ This could be due to
instrument failure. In this
case, contact your dealer or
Mitutoyo sales office.
Setting value rewrite error
・ A setting value has been
rewritten from any cause
after the previous use.
・ Press the center key will
return to the initial state
after setting the battery.
Perform re-setting after
confirming the setting
value, referring to "2.1
Installation (replacement) of
Battery and Initial Setting".
Setting value storage error
・ A setting value could not
be stored.
・ A setting value could not
be loaded.
・ Remove the battery once
and perform re-setting,
referring to "2.1 Installation
(replacement) of Battery
and Initial Setting".
・ If this error occurs even
after re-setting the battery,
replace it.
・ If this error occurs even
after replacing the battery, it
could be due to instrument
failure. In this case, contact
your dealer or Mitutoyo
sales office.