No. 99MAH040B
3-14
3.3.1 TOL: Tolerance judgment
The tolerance judgment function compares measurement data (display value) with
tolerance values (upper/lower limit) for OK/NG judgment. The tolerance values can be set
in preset No. (P1, P2, P3) respectively.
Enabling tolerance judgment. Changing tolerance values
1. Press the upper right key in the measurement mode to move the setup mode.
2. Press the lower left or lower right key to move the blinking cursor to "TOL".
3. Press the center key to move the ON/OFF selection of the tolerance judgment
function.
4. Press the lower left key to select "ON".
5. Press the center key to move the tolerance value setting (upper/lower limit).
The upper limit sign is blinking, and currently set value will be displayed.
6. When press the lower left key, the display switches to the lower limit. (Pressing the
lower left key, the upper limit and the lower limit switch.)
The lower limit sign (or the upper limit sign) is blinking, and the currently set lower
limit (or the upper limit) will be displayed.
7. To edit the tolerance value, select the target value and press the lower right key.
The numerical value editing will be enabled. (Refer to "3.2.2 Numerical value
editing".)
8. Press the center key after checking or editing the tolerance values. The tolerance
judgment is set "ON", and the mode returns to the parameter setting. (Refer to
"3.3 Setup Mode".)
Disabling tolerance judgment
1. Press the upper right key in the measurement mode to move the setup mode.
2. Press the lower left or lower right key to move the blinking cursor to "TOL".
3. Press the center key to move the ON/OFF selection of the tolerance judgment
function.
4. Press the lower left key to select "OFF".
5. Press the center key to set the tolerance judgment "OFF" and return to the
parameter setting. (Refer to "3.3 Setup Mode".)
IMPORTANT
The tolerance values will be set in the currently active preset No. (P1, P2, P3) .
Check the desired measuring system is selected before setting the tolerance judgment
function.
The tolerance values will be calculated automatically according to the unit or the
resolution. Check the tolerance values after the resolution is changed since the
conversion error may occur.
NOTE
Press the upper left key to cancel the operation.
If the upper limit is set lower than the lower limit, the tolerance upper/lower limit setting
error (Err 90) will occur. Reset the values to be the upper limit is greater than the lower
limit.
The tolerance judgment setting cannot be completed if Overflow error of upper / lower
limit value (Err95) is occurring. Reset the upper or lower limit value.
When the tolerance judgment setting is changed, all the measurement values saved
with the data save function will be deleted.