4-3
4.4 Timing Chart
IMPORTANT
If the output request (REQ) signal is received during the spindle displacement or if the
inverter receives output request (REQ) signals repeatedly at short intervals, the
instrument may not output data.
NOTE
Retain the REQ signal at Low level until the CK signal is outputted. Also, return the
REQ signal to High level before the last CK signal (at the 52nd bit) is outputted.
T5 T2 T3 T4T1
REQ
DATA
CK
RD
T5 T2 T3 T4T1
REQ
DATA
CK
RD
RD
REQ
CK
DATA
0.7 ms < T1 < 1.2 ms
100 µs < T2 < 200 µs
200 µs < T3 < 300 µs
100 µs < T4 < 200 µs
T5
Determined by the performance
of the data processor