No. 99MAH040B
3-2
3.1 Power ON/OFF
Power ON: Press the upper left key.
Power OFF: Press and hold the upper left key
NOTE
If the power is turned off right after the battery is installed before completing preset
setting, "------" will be displayed when the power is turned on again.
When the power is turned off while the instrument is holding the display value, and
then the power is turned on again, the held display value will be released.
When the power is turned off, all the measurement values saved with the data save
function will be deleted.
Auto-off function (energy saving function): The power automatically will be turned off after
approximately 20 minutes of idle time.
NOTE
When enabling FAST mode, auto-off function will be disabled. (Refer to "3.3.6.3
FAST(3): FAST mode ".)
3.2 Measurement mode
The following describes each setting and operation of the measurement mode of this
instrument. Refer to "1.1 Overview" for details of the measurement mode.
3.2.1 Preset setting
Set up the preset when performing master setting.
Three settings; P1, P2, P3 are available as preset values.
1. Starting the preset setting
Press and hold the lower left key in the measurement mode. "P" ( indicates
Preset No.) starts blinking and the previously set value will be displayed. If the value
does not need to be changed, skip to step 4.
2. Selecting preset No.
Press the lower left key and select the preset No. to be set. Pressing the lower left key,
the display changes in the following order: P1P2P3.
3. Editing the preset value.
Press the lower right key to move the numerical value editing. Refer to "3.2.3
Numerical value editing" for details of the setting.
4. Completing the preset setting
Lift up the spindle and position the contact point in the desired preset position. Next,
press the center key.
In the peak detection mode, the spindle position of minimum value will be set as the
preset position.