5-3
Display
definition
Actions
Internal program error
Measurement cannot be
performed due to
occurrence of an
anomaly in the internal
program.
This could be due to
instrument failure. In this
case, contact your dealer or
Mitutoyo sales office.
Tolerance upper/lower limit
setting error
The tolerance limit value
is set with the upper limit
value being smaller than
the lower limit value.
Press the center key to
reset to be the upper limit
value is greater than the
lower limit value. (Refer to
"3.3.1 TOL: Tolerance
judgment".)
Overflow error of preset
value
The preset value has
exceeded the maximum
number of digits that can
be displayed.
Press the lower right key to
reset the preset value.
Lower the resolution. (Refer
to "3.3.2 RES: Resolution".)
Overflow error of upper limit
value
The upper limit value has
exceeded the maximum
number of digits that can
be displayed.
Press the lower right key to
reset the upper limit value.
(Refer to "3.3.1 TOL:
Tolerance judgment".)
Lower the resolution. (Refer
to "3.3.2 RES: Resolution".)
Overflow error of lower limit
value
The lower limit value has
exceeded the maximum
number of digits that can
be displayed.
Press the lower right key to
reset the lower limit value.
(Refer to "3.3.1 TOL:
Tolerance judgment".)
Lower the resolution. (Refer
to "3.3.2 RES: Resolution".)