88
Alert:   Unable to enter KVO rate that is greater than infusion rate
Cause:   Occurs when user enters a KVO rate that is greater than infusion rate.
Resolution:  Recheck data entry. KVO rate must be less than programmed or calculated
  infusion rate.
LCD Display:   ***ALERT *** LESS THAN MIN FIELD LIMIT, MIN = ________ 
Alert:   Unable to program a field with a value that is less than the minimum limit.
Cause:   Value entered is less than the minimum limit for the pump or medLMTS
™ 
 programmed.
Resolution:  Recheck data entry. A value entered cannot be less than the field’s minimum  
 limit.
LCD Display:  ***ALERT *** “LOW C BATTERIES” 
Alert:   Battery source power is low.
Cause:   The batteries that are powering the pump have reached levels that may prevent
  the continued operation of the pump beyond 1 hour (actual time is rate 
  dependent). Alert applies only to C-cell batteries, not to the rechargeable 
  battery pack.
Resolution:  PAUSE the therapy in progress, turn the pump “OFF”; and then replace the 
  batteries. (See chapter 1 of the User’s Manual for instructions on battery 
  change.)  
LCD Display:  ***ALERT *** MAX CHANGES FOR PERIOD REACHED
Alert:   The maximum number of changes for the current delta time window has been
 reached.
Cause:  The maximum number of pump rate changes for the current delta time 
  window has been reached. This occurs when the rate has been changed more 
  than 12 times during any delta time window.
Resolution:  Allow more time to elapse before attempting to change the rate or reprogram 
  the therapy with medLMTS
™ and prescription basal and bolus ranges that 
will 
  reduce the number of times that the rate will have to be adjusted.
LCD Display:  ***ALERT *** MODIFICATIONS NOT ALLOWED
Alert:   Modification Not Allowed.
Cause:   The selection cannot be modified at this time or the programmed lock setting
  does not permit modifications to this selection.
Resolution:  Review the User’s Manual for program restrictions or if lock level restricted,
  change lock setting to access area you wish to modify.
LCD Display:   ***ALERT *** NEW DOSE IS LESS THAN CURRENT DOSE
Alert:   Unable to program a lower number of doses than number of doses already 
 infused.