SECTION 1
2
System Description
The X-Scribe is a high-performance, PC-based, dedicated color, stress exercise testing system capable of acquiring,
viewing, printing, and storing ECG, NIBP, and (optional) SpO
2
data. The X-Scribe may be used with a treadmill,
ergometer, or pharmacological form of testing. The device is equipped with Mortara Instrument’s VERITAS™
resting ECG interpretation algorithm. This feature enables the VERITAS algorithm to provide an over-reading
physician with a silent second opinion through diagnostic statements output on the ECG report. (The interpretive
portion of the ECG is available in the pre-exercise portion of the test only.) For additional information on the
VERITAS algorithm, please refer to the Physician’s Guide to Resting Interpretation. (See Parts and Accessories.)
The X-Scribe includes:
PC with keyboard and mouse
24” color monitor (1920 x 1080)
Z200+ thermal array printer
Isolation transformer
Patient cable with USB interface module
Patient lead wires
Patient cable belt
Software
System cart
Optional items include:
High-speed laser printer
Treadmill
Ergometer (Lode or Ergoline)
Multi-modality module for wireless USB 2500 MHz RF transmitter
Multi-modality module for wireless USB 608 MHz RF transmitter
Quick-connect module for use with either M12A or M12B front end
Programmable remote review workstation
Programmable remote keypad
Local area network (LAN)
NIBP with (optional) SpO
2
NIBP without SpO
2
Miscellaneous System Information
X-Scribe is capable of supporting the following video resolutions: 1280 x 1024 at 16M colors, 1920 x
1080, and 1920 x 1200.
X-Scribe is capable of supporting HP LaserJet printers with 600 dpi and PCL5 capabilities, and Mortara
Instrument’s Z200+ thermal printer.
The Network option is automatically disabled during the stress test; it will reinitiate at test completion.
A network isolation device supplied with the Network option must be installed in line between the PC and
wall network plug. This is to ensure compliance with IEC 60601-1-1 standard.
NOTE: No user-serviceable parts are inside. ANY modification to device may alter defibrillator
protection. Any modification to any part of device is to be performed by qualified service personnel only.