APPENDIX F
77
Maintaining the Z200+ Thermal Printer
If the hospital or institution fails to implement a satisfactory cleaning and inspection schedule for this equipment, it
may result in equipment failure and health hazards.
NOTE: Only qualified service personnel should repair or replace Z200+ thermal printer parts.
Inspect the equipment for the following conditions on a regular basis:
Examine power cable and the network cable for obvious damage (i.e., torn insulation, broken connectors,
etc.). Replace cables as necessary.
All cords and connectors are securely seated in their corresponding connections.
Examine the equipment for missing screws, cracks, or broken areas that might allow unintended access to
internal electronics areas.
Cleaning the Z200+ Thermal Printer
NOTE: If using a laser printer, refer to the printer user manual for maintenance and cleaning instructions.
To clean the printer
1. Disconnect the power source.
2. Clean the exterior surface of the unit with a damp cloth using a solution of mild dishwashing detergent
diluted in water.
3. After washing, thoroughly dry off the unit with a clean, soft cloth or paper towel.
To clean the print head
NOTE: Do not let soap or water come into contact with the writer, plugs, jacks, or vents.
1. Open writer door.
2. Lightly rub print head with an alcohol pad.
3. Wipe with a clean cloth to remove alcohol residue.
4. Allow print head to air dry.
5. Clean the platen by using adhesive tape. Apply the tape and pull it off. Rotate roller and repeat until entire
roller is clean.
6. Clean cue sensor photo detector.

Table of Contents