PROTOCOLS
APPENDIX B
57
The following protocols are delivered with every X-Scribe:
Bruce
Modified Bruce
Naughton
Modified Balke
High Ramp
Medium Ramp
Low Ramp
USAF/SAM 2.0
USAF/SAM 3.3
Astrand (Ergometer)
Pharmacological
Ellestad
Bicycle
These sample protocols produce the following operations and conditions:
An automatic blood pressure acquisition towards the end of each stage.
An automatic ECG report at the beginning of each stage.
In the Recovery phase, an automatic blood pressure measurement every 2 minutes and an ECG report
every minute until the end of the Recovery phase.
NOTE: Protocols are subject to physician preference and may be modified as deemed appropriate.
Bruce Protocol
The sample Bruce protocol produces the following operations and conditions:
A stage change occurs every 3 minutes with an increase in treadmill speed and grade.
An automatic blood pressure measurement is initiated one minute before the end of each stage.
A 12-lead ECG report is automatically printed at the end of each stage. ECG acquisition begins 12
seconds prior to the end of a stage.
In the Recovery phase, the treadmill slows to 0.5 mph and the following reports are automatically printed:
Peak Exercise ECG report prints immediately.
Recovery ECG prints in one minute.
Recovery ECG prints every one minute until the end of the Recovery phase.
Blood pressure measurement is set to every 2 minutes until the end of the Recovery phase.