CART WITH IV POLE
PARAMETER SPECIFICATION
HANGERS 2
HANGER CAPACITY 2 – 500cc glass bottles
POLE TRAVEL
64 to 90 inches (measured from the ground to the
POLE CONTROL
Foot pedal activation or BOTTLE soft-key
4.4 ENVIRONMENTAL REQUIREMENTS
The Model 3000 should be used in a clean environmental that is suitable for surgery.
The following environmental LIMITATIONS are as follows:
ENVIRONMENT OPERATING STORAGE/TRANSPORT
ALTITUDE ≤6,561 feet (2,000 meters) ≤14,000 feet (4,267 meters)
TEMPERATURE 50°F to 95°F (10°C to 35°C) 41°F to 140°F (5°C to 60°C)
RELATIVE HUMIDITY 30% to 85% RH 30% to 85% RH
ATMOSPHERIC
79 kPa - 102kPa ---
4.5 FACTORY DEFAULT SETTINGS
MODE
F
LOW
(cc/min)
VACUUM
(mmHg)
P
OWER
(%)
P
ULSE
(pulses/s)
FUNCTION
CUT RATE
(cuts/min)
BURST
(%- ms)
O
N
/O
FF
(ms)
IV
P
OLE
H
EIGHT
(in.)
IRR
77
U/S
Normal
24
200
40
0
Linear Power
77
U/S
Burst
24
200
45
Multi Burst
20
77
U/S
Micro
24
200
45
Linear Power
20 / 80
77
I/A
26
460
Linear Flow /
Fixed Vac
84
I/A
Cap
Vac
6
20
Linear Flow /
Fixed Vac
84
VIT
20
120
Fixed Vac /
200
74
COAG
40
Linear Power