7.8 SYSTEM MESSAGES
These series of messages are typically displayed during the operation of the system.
U/S PRIMING MESSAGES DESCRIPTION
Priming… Please Wait!
“PRIME” button pressed - will clear when finished
priming; automatically goes to tune handpiece.
Vac Test Failed! Prime Again!
Can’t build vacuum when testing during prime - check
the tubing set for open connections, leaks.
Fluid Path Blocked! Prime Again!
Vacuum too high during prime - check that the
administration tubing set roller clamp is released and
irrigation solution is flowing.
Check that the tubing set is not pinched or blocked.
OCCLUSION! Phaco tip occluded, vacuum reached maximum value.
Reflux Reflux activated on the foot pedal.
Handpiece Tuned
Message appears at the end of a prime cycle or
after “Tune” button is pressed.
Check Cassette!
Cassette not seated properly - should click into place
when fully seated across the locking mechanism.
Check Handpiece!
Handpiece not plugged in, tip not fully seated onto
handpiece and / or handpiece failed tuning.
Check Footpedal!
Check cable of foot pedal for condition,
Check foot pedal not plugged in.
Purging… Please Wait!
Purge button pressed - will clear after tubing is
purged.
I/A and VIT MESSAGES DESCRIPTION
OCCLUSION! Phaco tip occluded, vacuum reached maximum value.
Reflux Reflux activated on the foot pedal.
Handpiece Tuned
Message appears at the end of a prime cycle or after
“Tune” button is pressed.
Check Cassette!
Cassette not seated properly - should click into place
when fully seated across lock mechanism.
Check Footpedal!
Check cable of foot pedal for condition,
Check foot pedal not plugged in.
Pump stopped
In Linear Vacuum: when the vacuum limit is reached,
the peristaltic pump will stop.