8.0 SYSTEM CLEANING
This section of the manual contains the instructions for cleaning of the Model 3000 console,
foot pedal, cart, and accessories.
For cleaning and sterilization instructions of the following system components, refer to the
individual Instruction for Use (IFU) packaged with the item for explicit details on proper care and
handling, where applicable:
Ultrasonic Handpiece
Reusable Tubing Set
Phaco Tip (and Tip Wrench)
Silicone Irrigation Sleeve
Test Chamber
IV Pole
For cleaning and sterilization instructions of additional accessories included with your system
(optional items NOT MANUFACTURED BY Medical Technical Products, LP), the Instructions for
Use (IFU) are provided by the accessory manufacturer, where applicable:
Cautery
Vitrectomy
I/A Accessories
WARNING: Sterility assurance is the responsibility of the end user facility. Always follow
the
facility’s requirements for items that need to be cleaned and sterilized.
WARNING: Inadequate cleaning may result in debris exfoliating into the surgical field
.
WARNING: Clean console, foot pedal, cart and accessories immediately after every procedure
.
CAUTION: Use of cleaning solutions other than the ones listed in this section may damage the
surface of the console and void your warranty!
8.1 PURGING THE SYSTEM CONSOLE
8.1.1 Close the roller clamp on the administration tubing set.
8.1.2 Disconnect tubing kit from handpiece.
8.1.3 Connect the aspiration fitting of tubing kit (female fitting) to the irrigation fitting of
tubing kit (male fitting).
8.1.4 Verify that a collection bag is installed to the collection tube of the tubing kit.
8.1.5 On the screen of the console, select U/S mode.
8.1.6 On the screen of the console, select Purge.
8.1.7 Once the system has completed the purge cycle, on the back panel of the system,
press the power switch to the “O” OFF position.