10.0 SYSTEM TROUBLESHOOTING
AC POWER
PROBABLE CAUSE CORRECTIVE ACTION
1. AC Power loose or
disconnected
1. Reconnect AC power cord
1. AC line fuse blown 1. Replace fuses
PRIME CYCLE
PROBABLE CAUSE CORRECTIVE ACTION
“Check Cassette” Error Message
1. Priming cycle may have
caused cassette to become
dislodged
1. Remove and reinstall cassette
2. Run Prime cycle again
“Fluid path blocked, prime again” Error Message
1. Irrigation tubing may be
crimped, blocked or kinked
1. Free the irrigation tubing from occlusion and
run a prime cycle
2. Replace administration tubing set and run a
prime cycle
“Vac test failed, prime again” Error Message
1. Peristaltic tubing not
properly seated in groove in
cassette
1. Remove cassette and align peristaltic tubing in
groove
2. Fully seat cassette into locking mechanism and
run a prime cycle
2. Ultrasonic handpiece has a
leak
1. Remove irrigation and aspiration tubes from
the handpiece and connect them together.
2. Run a prime cycle. If prime cycle passes,
replace handpiece.
3. Defective tubing cassette 1. Replace tubing cassette and run a prime cycle.
4. Possible defective Model
3000
1. Contact your local distributor or MTP for
assistance