7.2 U/S PHACO MODE (SECTION 5.4)
7.2.1 Connect Ultrasonic Handpiece to Irrigation / Aspiration tubing.
7.2.1.1 Connect the aspiration tube from the cassette (female fitting) to the handpiece
aspiration port (male fitting).
7.2.1.2 Connect the irrigation tube from the cassette (male fitting) to the handpiece
irrigation port (female fitting).
7.2.1.3 Place a test chamber over the handpiece tip and sleeve until fully seated.
7.2.1.4 Depress foot pedal to position 1 to remove all air from tubing kit.
7.2.1.5 On the screen of the console, select Tune.
7.2.2 On the screen of the console, select U/S mode.
7.2.3 Select Multimode (1, 2, 3).
7.2.4 Select Phaco mode.
7.2.4.1 Normal.
7.2.4.1.1 U/S power delivery (Fixed, Linear).
7.2.4.1 Burst.
7.2.4.1.1 U/S power delivery (Single, Multi).
7.2.4.1 Micro.
7.2.4.1.1 U/S power delivery (Fixed, Linear).
7.2.5 Activate the ultrasonic function.
7.2.5.1 Depress foot pedal to your desired function of the U/S Phaco mode:
Position (1) for Irrigation only.
Position (2) for Irrigation and Aspiration only.
Position (3) for Irrigation, Aspiration, and U/S Phaco Power
7.2.5.2 While in any position on the foot pedal, make any necessary adjustments to
applicable parameters as needed by pressing the UP or DOWN arrows:
7.2.5.3 Fully release foot pedal to end irrigation, aspiration and ultrasonic function.