5.5.2 MULTIMODE SOFT-KEYS I/A1, I/A2 & CAP VAC
The Multimode soft-keys are used as I/A Mode presets for access to user-programmed I/A Modes.
Cap Vac is used to quickly change the Aspiration settings, when the doctor is polishing the
posterior capsule and requires very low Flow and Vacuum levels.
The user can select any of the three (3) multimode soft-keys and edit default I/A programs by
completing the following steps:
5.5.2.1 Select any multimode soft-key.
5.2.2 Edit any of the following I/A parameters for the multimode soft-key selected:
Flow (cc/min)
Fixed Flow
Linear Flow
Vacuum (mmHg)
Fixed Vac
Linear Vac
IV Pole Height
5.5.2.3 Select another multimode soft-key.
5.5.2.4 Edit any of the I/A parameters listed above in Section 5.5.2.2 for the multimode soft-
key selected:
5.5.2.5 Now select the first multimode soft-key you previously edited. The edited I/A
parameters should be as first edited.
NOTE: The edited I/A Mode parameters will remain until the system is
powered OFF.
I/A Mode parameters will return to default factory settings when the system is powered ON.