7.0 SYSTEM OPERATION
Before operating in any mode, verify that all accessories used during operation have been setup per
Section 6.2 Accessory Setup and verify that all per-operational checks have been performed per
Section 6.3 System Preoperational Checks.
If you want to use a program, press the program arrow and select your program.
Any necessary adjustments to parameters can be made by pressing the UP or DOWN arrow soft-keys.
Activating the assigned reflux side kick lever will enable the reflux function in any foot pedal position.
7.1 IRR MODE (SECTION 5.3)
7.1.1 Connect Handpiece to Irrigation / Aspiration tubing.
7.1.1.1 Connect the aspiration tube from the cassette (female fitting) to the handpiece
aspiration port (male fitting).
7.1.1.2 Connect the irrigation tube from the cassette (male fitting) to the handpiece
irrigation port (female fitting).
7.1.1.3 Place a test chamber over the handpiece tip and sleeve until settled firmly.
7.1.1.4 Fully depress foot pedal to remove all air from tubing kit.
7.1.1.5 Remove test chamber from the tip of the handpiece.
7.1.2 On the screen of the console, select IRR mode.
7.1.3 Activate the irrigation function.
7.1.3.1 Depress foot pedal to position 1 to begin irrigation.
7.1.3.2 While in any position on the foot pedal, make any necessary adjustments to the IV
pole height as needed by pressing the arrows:
7.1.3.3 Fully release foot pedal to end irrigation function.