What type of tubing should I use for “IV Bottle to Patient Height” with MTP 3000?
- FfritzcynthiaSep 16, 2025
If the MTP Medical Equipment is showing “IV Bottle to Patient Height” issues, use a high-flow, non-filtered IV administration tubing set.
What type of tubing should I use for “IV Bottle to Patient Height” with MTP 3000?
If the MTP Medical Equipment is showing “IV Bottle to Patient Height” issues, use a high-flow, non-filtered IV administration tubing set.
What to do if MTP 3000 Medical Equipment displays “Vac test failed, prime again” error?
If your MTP Medical Equipment shows a “Vac test failed, prime again” error message, it might be because the peristaltic tubing isn't properly seated in the groove in the cassette. Remove the cassette and make sure the peristaltic tubing is aligned correctly in the groove. Then, fully seat the cassette into the locking mechanism and run a prime cycle.
What to do if MTP Medical Equipment displays “Check Cassette” Error Message?
If the MTP Medical Equipment displays a “Check Cassette” error message, the priming cycle may have caused the cassette to become dislodged. Remove and reinstall the cassette, then run the Prime cycle again.
Why is there “No Irrigation Flow” in my MTP 3000?
If the MTP Medical Equipment is showing “No Irrigation Flow”, ensure that the roller clamp on the administration tubing is released to allow flow.
How to troubleshoot “Poor or No Ultrasonic (U/S) Power” on MTP 3000 Medical Equipment?
If you're experiencing “Poor or No Ultrasonic (U/S) Power” with your MTP Medical Equipment, the ultrasonic handpiece may be defective. Replace the ultrasonic handpiece and retune the system.
What if MTP 3000 Medical Equipment shows “Ultrasonic (U/S) Handpiece will not Tune” because of damaged phaco tip?
If the MTP Medical Equipment shows “Ultrasonic (U/S) Handpiece will not Tune”, the phaco tip might be damaged. Replace the damaged tip with a new phaco tip and retune the handpiece.
Why MTP 3000 shows “Ultrasonic (U/S) Handpiece will not Tune”?
If the MTP Medical Equipment shows “Ultrasonic (U/S) Handpiece will not Tune”, the phaco tip might be loose. Tighten the phaco tip with a tip wrench and retune the handpiece.
How to resolve “Vacuum ERROR Message” caused by closed roller clamp on MTP 3000?
If the MTP Medical Equipment displays a “Vacuum ERROR Message”, it could be due to a closed roller clamp on the irrigation tubing. Release the roller clamp on the irrigation line and verify flow.
How to adjust “IV Bottle to Patient Height” on MTP Medical Equipment?
If the MTP Medical Equipment is showing “IV Bottle to Patient Height” issues, it's possible the IV solution bottle is too low. Raise the solution bottle until the anterior chamber stabilizes. Ensure the patient’s eye level is appropriately positioned relative to the tubing cassette.
What to do if MTP 3000 Medical Equipment shows “No Irrigation Flow”?
If the MTP Medical Equipment is showing “No Irrigation Flow”, and the irrigation valve is closed, depress the foot pedal to verify the irrigation pinch valve “clicks”. If it doesn’t, remove and reseat the cassette.
Brand | MTP |
---|---|
Model | 3000 |
Category | Medical Equipment |
Language | English |
Covers critical safety warnings, electrical hazards, and usage precautions for the system.
Details unpacking, manual, console, handpiece, tubing, foot pedal, and optional cart.
Lists and describes console controls, indicators, and connections.
Explains various safety and operational symbols used throughout the manual.
Provides detailed technical specifications for the console, electrical, and ultrasonic functions.
Outlines environmental limitations and factory default operational settings for modes.
Explains the system software structure, main menu navigation, and operational modes.
Guides users through the main menu, operational modes, and soft-key functions.
Details the operation, control, and fluid delivery of the Irrigation mode.
Explains the Phaco mode, its functions, and different modes like Normal, Burst, Micro.
Covers the I/A mode, its functions, and control of irrigation and aspiration.
Describes the Vitrectomy mode, its functions, and foot pedal control for cutting.
Explains the Coagulation mode used for cauterizing small blood vessels.
Describes how to access and use the Help mode for on-screen assistance.
Details how to manage programs, settings, sound, voice, video, and preferences.
Explains foot pedal settings, calibration procedures, and side kick functions.
Step-by-step guide for recalibrating the foot pedal to desired positions.
Step-by-step instructions for setting up the console, power, and accessories.
Guides on setting up irrigation/aspiration, vitrectomy, cautery, and ultrasonic handpieces.
Details the Prime Cycle and System Fluidics Test for pre-use verification.
Details operations for IRR, U/S Phaco, I/A, VIT, and COAG modes.
Provides instructions for replacing irrigation solution container and ultrasonic handpiece.
Lists typical system messages and their meanings for various modes and functions.
Details cleaning procedures for the system console, foot pedal, and optional cart.
Provides instructions for cleaning accessories and system sterilization methods.
Step-by-step guide for replacing the AC power input module fuses.
Information on how to contact regional distributors for technical support.
Addresses troubleshooting for AC power problems and prime cycle errors.
Covers troubleshooting for irrigation flow, venting, and aspiration problems.
Addresses issues related to U/S power, tuning, vacuum errors, and loose tips.
Troubleshoots issues with Vitrectomy handpiece function and Coagulation power.
Addresses troubleshooting for IV bottle height and anterior chamber stability issues.
Presents power output data, impedance curves, and EMC compliance information.