MODEL 3000
PHACOEMULSIFICATION SYSTEM
Operator’s
Manual

Table of Contents

Questions and Answers

  • L
    Laura HigginsSep 12, 2025
    Why MTP 3000 shows “Ultrasonic (U/S) Handpiece will not Tune”?
    • P
      Phillip JonesSep 12, 2025
      If the MTP Medical Equipment shows “Ultrasonic (U/S) Handpiece will not Tune”, the phaco tip might be loose. Tighten the phaco tip with a tip wrench and retune the handpiece.
  • P
    pscottSep 9, 2025
    How to fix “No Irrigation Flow” due to improperly installed irrigation tubing in MTP 3000 Medical Equipment?
    • S
      Sheila HallSep 10, 2025
      If the MTP Medical Equipment is showing “No Irrigation Flow”, the irrigation tubing might be improperly installed. Disconnect the irrigation tubing and reinstall it correctly.
  • D
    dylan46Sep 8, 2025
    What if there is “No or poor aspiration” vacuum and tubing set is incorrectly installed in MTP Medical Equipment?
    • A
      Anne HaynesSep 8, 2025
      If you're getting “No or poor aspiration” vacuum on your MTP Medical Equipment, the tubing set might be incorrectly installed. Check the tubing set installation and correct the connections. Then, run a prime cycle to verify aspiration with NO Error message.
  • M
    Michael JohnsonSep 7, 2025
    Why is there “Poor or No Ultrasonic (U/S) Power” in MTP 3000?
    • M
      mercadotonySep 7, 2025
      If you're experiencing “Poor or No Ultrasonic (U/S) Power” with your MTP Medical Equipment, the foot pedal may not be fully depressed in Linear Power mode. Depress the foot pedal until power is displayed on the U/S Phaco screen.
  • G
    glenncrystalAug 23, 2025
    What to do if MTP 3000 shows “Fluid path blocked, prime again” error?
    • D
      Danielle HuynhAug 23, 2025
      If the MTP Medical Equipment displays a “Fluid path blocked, prime again” error message, it could be due to crimped, blocked, or kinked irrigation tubing. Try freeing the irrigation tubing from any obstructions and running a prime cycle again. If that doesn't work, replace the administration tubing set and run another prime cycle.
  • B
    Brent WintersSep 5, 2025
    What if there is “No Irrigation Flow” because of irrigation tube in MTP 3000?
    • O
      obrownSep 5, 2025
      If the MTP Medical Equipment is showing “No Irrigation Flow”, it could be due to a kinked or damaged irrigation tube. Try unkinking the tube, or replace it if needed.
  • M
    Manuel SavageSep 4, 2025
    How to adjust “IV Bottle to Patient Height” on MTP Medical Equipment?
    • K
      Kevin BurnsSep 4, 2025
      If the MTP Medical Equipment is showing “IV Bottle to Patient Height” issues, it's possible the IV solution bottle is too low. Raise the solution bottle until the anterior chamber stabilizes. Ensure the patient’s eye level is appropriately positioned relative to the tubing cassette.
  • N
    nbensonSep 3, 2025
    How to fix “No or poor aspiration” vacuum on MTP 3000?
    • C
      Clifford RodriguezSep 3, 2025
      If you're getting “No or poor aspiration” vacuum on your MTP Medical Equipment, the vacuum setting might be too low. Increase the vacuum and rerun the mode to clear the error message.
  • P
    palmershaunSep 2, 2025
    What to do if MTP 3000 Medical Equipment shows “No Irrigation Flow”?
    • W
      woodshawnSep 2, 2025
      If the MTP Medical Equipment is showing “No Irrigation Flow”, and the irrigation valve is closed, depress the foot pedal to verify the irrigation pinch valve “clicks”. If it doesn’t, remove and reseat the cassette.
  • M
    Mark VasquezAug 31, 2025
    What to do if MTP 3000 shows “Ultrasonic (U/S) Handpiece will not Tune”?
    • R
      Robert GonzalezSep 1, 2025
      If the MTP Medical Equipment shows “Ultrasonic (U/S) Handpiece will not Tune”, the ultrasonic handpiece may be defective. Replace the ultrasonic handpiece and retune to the system.

Summary

1.0 WARNINGS AND PRECAUTIONS

General Safety Precautions

Covers critical safety warnings, electrical hazards, and usage precautions for the system.

2.0 WARRANTY INFORMATION

3.0 PRODUCT INFORMATION

System Components and Setup

Details unpacking, manual, console, handpiece, tubing, foot pedal, and optional cart.

4.0 SYSTEM DESCRIPTION

System Overview and Controls

Lists and describes console controls, indicators, and connections.

System Signs and Symbols

Explains various safety and operational symbols used throughout the manual.

Technical Specifications

Provides detailed technical specifications for the console, electrical, and ultrasonic functions.

Environmental and Default Settings

Outlines environmental limitations and factory default operational settings for modes.

5.0 SYSTEM OPERATION OVERVIEW

System Software Overview

Explains the system software structure, main menu navigation, and operational modes.

Main Menu Navigation

Guides users through the main menu, operational modes, and soft-key functions.

Irrigation (IRR) Mode

Details the operation, control, and fluid delivery of the Irrigation mode.

Phaco (U;S) Mode

Explains the Phaco mode, its functions, and different modes like Normal, Burst, Micro.

Irrigation;Aspiration (I;A) Mode

Covers the I/A mode, its functions, and control of irrigation and aspiration.

Vitrectomy (VIT) Mode

Describes the Vitrectomy mode, its functions, and foot pedal control for cutting.

Coagulation (COAG) Mode

Explains the Coagulation mode used for cauterizing small blood vessels.

Help Mode

Describes how to access and use the Help mode for on-screen assistance.

Options Mode

Details how to manage programs, settings, sound, voice, video, and preferences.

Foot Pedal Mode

Explains foot pedal settings, calibration procedures, and side kick functions.

Foot Pedal Calibration Procedure

Step-by-step guide for recalibrating the foot pedal to desired positions.

6.0 SYSTEM SETUP AND PREOPERATIONAL CHECKS

System Setup Procedures

Step-by-step instructions for setting up the console, power, and accessories.

Accessory Setup Instructions

Guides on setting up irrigation/aspiration, vitrectomy, cautery, and ultrasonic handpieces.

Preoperational Checks

Details the Prime Cycle and System Fluidics Test for pre-use verification.

7.0 SYSTEM OPERATION

Mode Operations

Details operations for IRR, U/S Phaco, I/A, VIT, and COAG modes.

Replacing System Components

Provides instructions for replacing irrigation solution container and ultrasonic handpiece.

System Messages

Lists typical system messages and their meanings for various modes and functions.

8.0 SYSTEM CLEANING

Console, Foot Pedal, and Cart Cleaning

Details cleaning procedures for the system console, foot pedal, and optional cart.

Accessory Cleaning and Sterilization

Provides instructions for cleaning accessories and system sterilization methods.

9.0 SYSTEM SERVICE

Fuse Replacement

Step-by-step guide for replacing the AC power input module fuses.

Technical Support

Information on how to contact regional distributors for technical support.

10.0 SYSTEM TROUBLESHOOTING

AC Power and Prime Cycle Troubleshooting

Addresses troubleshooting for AC power problems and prime cycle errors.

Irrigation and Aspiration System Issues

Covers troubleshooting for irrigation flow, venting, and aspiration problems.

Phaco Mode Troubleshooting

Addresses issues related to U/S power, tuning, vacuum errors, and loose tips.

Vitrectomy and Coagulation Troubleshooting

Troubleshoots issues with Vitrectomy handpiece function and Coagulation power.

Anterior Chamber Stability Issues

Addresses troubleshooting for IV bottle height and anterior chamber stability issues.

11.0 SYSTEM PERFORMANCE

Performance Data and EMC Requirements

Presents power output data, impedance curves, and EMC compliance information.

MTP 3000 Specifications

General IconGeneral
BrandMTP
Model3000
CategoryMedical Equipment
LanguageEnglish