MODEL 3000
PHACOEMULSIFICATION SYSTEM
Operator’s
Manual

Table of Contents

Questions and Answers

  • K
    Katherine HenryAug 30, 2025
    What causes “No or poor aspiration” vacuum in MTP Medical Equipment?
    • M
      Melanie GoodmanAug 30, 2025
      If you're getting “No or poor aspiration” vacuum on your MTP Medical Equipment, make sure the system is in I/A Mode.
  • A
    Austin AtkinsonAug 29, 2025
    Why is there “NO Coag Power” on my MTP 3000?
    • N
      nbrooksAug 29, 2025
      If there is “No Coag Power” in your MTP Medical Equipment, the system might be in the wrong mode. Select the COAG Mode and rerun the procedure.
  • L
    Lauren JacksonAug 28, 2025
    What causes “Vacuum ERROR Message” on MTP 3000?
    • D
      David DelgadoAug 28, 2025
      If the MTP Medical Equipment displays a “Vacuum ERROR Message”, it could be due to vacuum present in the tubing. Disconnect and reconnect the tubing to relieve the vacuum in the lines.
  • S
    Sabrina MendozaAug 27, 2025
    How to troubleshoot “Poor or No Ultrasonic (U/S) Power” on MTP 3000 Medical Equipment?
    • C
      Christian BernardAug 27, 2025
      If you're experiencing “Poor or No Ultrasonic (U/S) Power” with your MTP Medical Equipment, the ultrasonic handpiece may be defective. Replace the ultrasonic handpiece and retune the system.
  • A
    Amber PerryAug 27, 2025
    Why is there “No Irrigation Flow” in my MTP 3000?
    • V
      villanuevatammyAug 27, 2025
      If the MTP Medical Equipment is showing “No Irrigation Flow”, ensure that the roller clamp on the administration tubing is released to allow flow.
  • M
    Melissa BakerAug 25, 2025
    What to do if MTP Medical Equipment displays “Check Cassette” Error Message?
    • H
      hbrockAug 26, 2025
      If the MTP Medical Equipment displays a “Check Cassette” error message, the priming cycle may have caused the cassette to become dislodged. Remove and reinstall the cassette, then run the Prime cycle again.
  • R
    rharmonAug 25, 2025
    What to do if MTP 3000 Medical Equipment displays “Vac test failed, prime again” error?
    • M
      monica73Aug 25, 2025
      If your MTP Medical Equipment shows a “Vac test failed, prime again” error message, it might be because the peristaltic tubing isn't properly seated in the groove in the cassette. Remove the cassette and make sure the peristaltic tubing is aligned correctly in the groove. Then, fully seat the cassette into the locking mechanism and run a prime cycle.

Summary

1.0 WARNINGS AND PRECAUTIONS

General Safety Precautions

Covers critical safety warnings, electrical hazards, and usage precautions for the system.

2.0 WARRANTY INFORMATION

3.0 PRODUCT INFORMATION

System Components and Setup

Details unpacking, manual, console, handpiece, tubing, foot pedal, and optional cart.

4.0 SYSTEM DESCRIPTION

System Overview and Controls

Lists and describes console controls, indicators, and connections.

System Signs and Symbols

Explains various safety and operational symbols used throughout the manual.

Technical Specifications

Provides detailed technical specifications for the console, electrical, and ultrasonic functions.

Environmental and Default Settings

Outlines environmental limitations and factory default operational settings for modes.

5.0 SYSTEM OPERATION OVERVIEW

System Software Overview

Explains the system software structure, main menu navigation, and operational modes.

Main Menu Navigation

Guides users through the main menu, operational modes, and soft-key functions.

Irrigation (IRR) Mode

Details the operation, control, and fluid delivery of the Irrigation mode.

Phaco (U;S) Mode

Explains the Phaco mode, its functions, and different modes like Normal, Burst, Micro.

Irrigation;Aspiration (I;A) Mode

Covers the I/A mode, its functions, and control of irrigation and aspiration.

Vitrectomy (VIT) Mode

Describes the Vitrectomy mode, its functions, and foot pedal control for cutting.

Coagulation (COAG) Mode

Explains the Coagulation mode used for cauterizing small blood vessels.

Help Mode

Describes how to access and use the Help mode for on-screen assistance.

Options Mode

Details how to manage programs, settings, sound, voice, video, and preferences.

Foot Pedal Mode

Explains foot pedal settings, calibration procedures, and side kick functions.

Foot Pedal Calibration Procedure

Step-by-step guide for recalibrating the foot pedal to desired positions.

6.0 SYSTEM SETUP AND PREOPERATIONAL CHECKS

System Setup Procedures

Step-by-step instructions for setting up the console, power, and accessories.

Accessory Setup Instructions

Guides on setting up irrigation/aspiration, vitrectomy, cautery, and ultrasonic handpieces.

Preoperational Checks

Details the Prime Cycle and System Fluidics Test for pre-use verification.

7.0 SYSTEM OPERATION

Mode Operations

Details operations for IRR, U/S Phaco, I/A, VIT, and COAG modes.

Replacing System Components

Provides instructions for replacing irrigation solution container and ultrasonic handpiece.

System Messages

Lists typical system messages and their meanings for various modes and functions.

8.0 SYSTEM CLEANING

Console, Foot Pedal, and Cart Cleaning

Details cleaning procedures for the system console, foot pedal, and optional cart.

Accessory Cleaning and Sterilization

Provides instructions for cleaning accessories and system sterilization methods.

9.0 SYSTEM SERVICE

Fuse Replacement

Step-by-step guide for replacing the AC power input module fuses.

Technical Support

Information on how to contact regional distributors for technical support.

10.0 SYSTEM TROUBLESHOOTING

AC Power and Prime Cycle Troubleshooting

Addresses troubleshooting for AC power problems and prime cycle errors.

Irrigation and Aspiration System Issues

Covers troubleshooting for irrigation flow, venting, and aspiration problems.

Phaco Mode Troubleshooting

Addresses issues related to U/S power, tuning, vacuum errors, and loose tips.

Vitrectomy and Coagulation Troubleshooting

Troubleshoots issues with Vitrectomy handpiece function and Coagulation power.

Anterior Chamber Stability Issues

Addresses troubleshooting for IV bottle height and anterior chamber stability issues.

11.0 SYSTEM PERFORMANCE

Performance Data and EMC Requirements

Presents power output data, impedance curves, and EMC compliance information.

MTP 3000 Specifications