EasyManua.ls Logo

MTP 3000 - Create and Savea Custom Program

MTP 3000
93 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 44 of 90
5.9.2 CREATE AND SAVE A CUSTOM PROGRAM
5.9.2.1 Select and change the desired parameter settings in each mode (IRR, U/S, I/A, VIT,
COAG)
5.9.2.2 Make a list and write down each Mode and the values of each change for custom
program memory verification later in this procedure.
5.9.2.3 After all desired parameter settings and Mode functions have been changed, press the
“OPTIONS” soft-key.
5.9.2.4 Press the “Programs” tab.
5.9.2.5 Press the “Add” function soft-key.
5.9.2.6 Use the keyboard on the Programs tab to enter the 11 character custom program
name.
5.9.2.7 Press the “OK” function soft-key or the keyboard ENTER (
) key to save the custom
program, or press “Cancel” to exit and not save the custom program.
NOTE: The remaining steps in this procedure are for creating the first custom program on the
system. These steps may also have to be repeated if program memory is questionable.
5.9.2.8 After saving the custom program, verify the custom program memory:
5.9.2.8.1 Press the power button to the “OFF” position on the back of the console.
5.9.2.8.2 Wait for 10 seconds and then press the power button to the “ON” position.
5.9.2.8.3 Wait for the system to boot up to the “FACTORY” U/S Phaco mode screen.
5.9.2.8.4 Press thePROGRAM” DOWN arrow soft- key at the top of the screen, to the right
of the word FACTORY.
5.9.2.8.5 Select the custom program name and verify that all the displayed parameter
settings and mode functions match the list created earlier in this procedure.

Table of Contents