8.4 CLEANING THE OPTIONAL CART WITH IV POLE
8.4.1 On the back panel of the system, press the power switch to the O” OFF position.
8.4.2 Disconnect the Model 3000 from all electrical power.
8.4.3 Clean all surfaces of the cart with IV pole with either 70% isopropyl alcohol or a mild
soap and water solution.
8.4.4 Thoroughly wipe the cart with IV pole with a soft lint-free cloth to remove all cleaning
solution or soap and water. DO NOT leave any cleaning solution residue or water
spots on the cart with IV pole surfaces.
8.4.5 Gently clean the connecting cable at end of each day using either 70% isopropyl
alcohol or a mild soap and water solution.
8.4.6 After cleaning, examine the cart with IV pole connecting cable for nicks, tears and
wear fractures in the cable sheath.
NOTE: The cart with IV pole should be wiped clean of any fluids that come in contact with it
during the course of the surgery. Surfaces could corrode or rust if not thoroughly cleaned.
8.5 CLEANING ACCESSORIES
8.5.1 Immediately prior to initial and subsequent use, carefully disassemble all items and
soak in enzymatic detergent solution.
8.5.2 Connect a syringe (not supplied) to the accessory to be washed.
8.5.3 Draw enzymatic detergent into the syringe.
8.5.4 Disconnect syringe and expel solution. Repeat from 8.5.1 for each fluid pathway.
8.5.5 Soak in distilled / deionized water.
8.5.6 Connect syringe to the accessory to be flushed.
8.5.7 Draw distilled / deionized water into the syringe.
8.5.8 Disconnect syringe and expel solution. Repeat from 8.5.5 for each fluid pathway.
8.5.9 To rinse the accessory, fill syringe with distilled / deionized water.
8.5.10 Expel the distilled / deionized water.
8.5.11 Ensure the free flow of distilled / deionized water occurs when expelling.
Repeat 2X from 8.5.9 for each fluid pathway.
8.5.12 To dry the accessory, expel an empty syringe through the accessory.
Repeat for each fluid pathway.