5.9.17 FOOT PEDAL CALIBRATION PROCEDURE (CONTINUED)
Remember, you have three (3) foot pedal positions. Wherever you choose to press and hold the
main treadle for each position is up to you. There is no set standard.
5.9.17.6 Follow the directions on the screen to calibrate the foot pedal.
5.9.17.6.1 Depress and hold the foot pedal at the “first” detent position and press the “OK”
soft-key.
5.9.17.6.2 Depress and hold the foot pedal at thesecond detent position and press the
“OK” soft-key.
5.9.17.6.3 Completely depress and hold the foot pedal to the “third” detent position and
press the “OK” soft-key.
5.9.17.6.4 Release the foot pedal and press theOK” soft-key.
NOTE: If the Foot Pedal Vibration is “ON”, the vibration will occur between positions
(1) and (2) and between positions (2) and (3).