Configuration
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch 101/144
117
11.1.2 Performing a firmware update with EasyOne Air
>
Required material
An SD memory card from the manufacturer
In this menu, you can update the firmware of EasyOne Air.
CAUTION
FAILED UPDATE
An update may fail unnoticed.
After performing an update, restart EasyOne Air and EasyOne Connect and per-
form a calibration check.
Calibration check, 118
Checking for correct operation of EasyOne Air, 122
1 Unlock and open the battery compartment by turning the lock screw
clockwise.
2 Insert the SD memory card into the SD card slot.