Configuration
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch 117/144
117
11.2.9 Description of environment configurations of EasyOne Connect
Utilities > Configuration > Environment
The environment conditions are required to accurately calculate the
diagnostic parameters from the raw sensor data.
If you select the check box, you are prompted every two hours to enter the
environment conditions when performing spirometry.
From the environment conditions, the ATPS values (Ambient Temperature
Pressure Saturated), which describe the conditions for the spirometer, are
calculated. The ATPS values are converted to BTPS values (Body Temperature
Pressure Saturated), which describe the body conditions of the patient.
If you choose one of the following tests, you are prompted for the current
environment conditions:
FVL test
MVV test
SVC test
Provocation test (if the FVL test is configured for the provocation test)