Safety information
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch 11/144
16
2 Safety information
2.1 Classification
In this manual, safety information is classified as follows:
WARNING
WARNING …
… indicates a hazard. If not avoided, the hazard can result in death or serious
injury.
CAUTION
CAUTION …
… indicates a potential hazard. If not avoided, the hazard may result in minor
injury and/or product/property damage.
2.2 General safety information
Electric Shock Patients and technicians may be exposed to dangerous voltage.
Connect only medical-grade power supply released by the manufacturer.
To ensure electric safety, connect only equipment, such as printers and net-
works, that complies with the IEC 60950-1 standard for the electrical safety of
IT equipment.
Patient health
hazard
Inadequate qualification of personnel who operate EasyOne Air can put the
patient’s health at risk.
The patient may suffer an asthma attack or a bronchospasm while performing
spirometry.
EasyOne Air must only be used by qualified personnel.
Do not leave the patient unattended while performing spirometry or in be-
tween trials.
Intended audience of this Operator’s Manual, 3
Cross-
contamination and
adverse reaction
After use, the disposable EasyOne Flow Tube can be infectious.
Provide means for proper disposal of infectious waste, according to legal re-
quirements.
False diagnosis Inadequate qualification of the personnel that operates EasyOne Air can lead
to false results, false interpretation, and false diagnosis.
EasyOne Air must only be used by qualified personnel.
Intended audience of this Operator’s Manual, 3
Electric shock Do not open EasyOne Air.