Safety information
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch 12/144
16
Fire Follow the manufacturer’s recommendation for proper disposal of the battery.
Only use an original battery by the manufacturer.
Do not store EasyOne Air close to flammable materials like paper or textiles.
Disposal, 4
List of order numbers and accessories for EasyOne Air, 131
False diagnosis EasyOne Air may be damaged when dropped or during transport.
Perform an incoming inspection when you first receive EasyOne Air. If you de-
tect any damage, contact your EasyOne Air dealer or the ndd Servicing Depart-
ment.
Especially check the battery for visible damage before use.
Do not drop EasyOne Air.
If EasyOne Air has been dropped, check for correct operation of EasyOne Air.
Checking for correct operation of EasyOne Air, 122
Fire due to explosive
or flammable gases
If used in the vicinity of flammable gases, the plastic case may catch fire and
cause burn injuries to the patient.
For preventive measures, make sure that the device is not used near flamma-
ble gases (for example, anesthetic agents).
If not in daily use, remove the battery from EasyOne Air and store the battery
in the original packaging.
False diagnosis Environment conditions outside the range described in the specifications can
cause measurement errors.
Use EasyOne Air only within the defined environment specifications for tem-
perature, humidity, and atmospheric pressure/altitude.
Observe the limits of storage conditions.
Setting up the EasyOne Air spirometer, 24
List of specifications for EasyOne Air, 125
Configuration of EasyOne Air, 93
List of specifications for EasyOne Air, 125
About reactivating EasyOne Air after storage, 124
False diagnosis Setup with wrong environment data can cause false results.
Verify the environment data that you have set up.
False diagnosis Electromagnetic fields from other devices may cause disturbances in
EasyOne Air.
False diagnosis Graphs on the display of EasyOne Connect may be misleading due to unclear
curve scaling between different patients.
Only conduct result interpretation of curves with a printed report.
About interpreting results, 72
List of quality messages and quality grades, 73