Safety information
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch 13/144
16
Failed update An update may fail.
Perform backups of the database of EasyOne Air and EasyOne Connect fre-
quently.
Performing a backup of EasyOne Air, 102
Data loss Data loss may occur for various other reasons and is generally unpredictable.
Backup the database of EasyOne Air frequently.
Performing a backup of EasyOne Air, 102
False diagnosis For unforeseen reasons, malfunction of EasyOne Air can lead to false results.
Perform calibration checks periodically.
Calibration check, 118
Checking for correct operation of EasyOne Air, 122
Malfunction Viruses, malware, and other hazardous software on your PC may adversely af-
fect the performance of EasyOne Connect.
Install anti-virus software on your PC.
Malfunction Damaged components can lead to false results.
Make sure that no liquid penetrates into EasyOne Air, for example through the
battery compartment or the USB port.
Leakage of acids
and explosion
Do not connect the battery improperly.
Do not throw the battery into fire.
Remove the battery from the battery compartment if you do not intend to use
EasyOne Air for a longer period.
Avoid mechanical shock to the battery.
If the rechargeable battery is damaged, liquids from the battery must not get
into contact with skin or eyes. In case of contact with skin or eyes, rinse thor-
oughly with water and consult a physician.
Only charge the battery using EasyOne Air, the power supply released by the
manufacturer, and the cradle.
Follow the manufacturer’s recommendation for proper disposal of the battery.
About charging the battery of EasyOne Air, 122
Disposal, 4
Malfunction Servicing must only be carried out by ndd staff or by qualified service person-
nel of official ndd distributors.
Do not open EasyOne Air.
Malfunction Non-original accessories and disposables, like respiratory tubes, battery,
power adapter, or cables can cause malfunction.
Only use original accessories and disposables by the manufacturer.
List of order numbers and accessories for EasyOne Air, 131
False diagnosis Non-original respiratory tubes can cause measurement error and false results.
You must only use EasyOne Flow Tube respiratory tubes by the manufacturer
ndd to assure accuracy, long-life, and full warranty coverage.