Configuration
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch 94/144
117
11.1.1.2 Description of spirometry settings of EasyOne Air
> >
Predicted normal values setting
Predicted normal values are based on the latest scientific publications.
Therefore, the predicted normal values are subject to change. These changes
are delivered to you via software updates.
Updated information is published in an Application Note on the ndd website.
Application Note Reference Predicted Normal Values
Application Notes for further information, 3
List of bibliographic references, 132
Best trial/best value selection
If you choose the Best Trial option, the complete best trial is selected for
assessing the quality of the test and for assessing the interpretation. The trial
with the largest sum of FVC and FEV
1
is selected, as suggested by ATS and
ERS.
If you choose the Best Value option, the best single, relevant values are
selected from different trials for assessing the quality of the test and for
assessing the interpretation. This compiled column shows, for example, the
largest FVC (or FEV
6
) and the largest FEV
1
from all acceptable tests (unless all
tests are unacceptable). Other parameters are taken from the best trial (also
defined by the largest sum of FEV
1
and FVC).