Hygiene and cleaning
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch 81/144
81
5 WARNING! Patient cross-contamination and infection of the technician:
Always wear disposable gloves. Between patients, always exchange
disposable gloves, clean EasyOne Air, and disinfect hands. / Electric
shock, failure of EasyOne Air, and false diagnosis: Make sure that no fluid
penetrates the EasyOne Flow Tube holder or the inside of EasyOne Air while
cleaning. To clean EasyOne Air, do not fully dip EasyOne Air into any fluid.
To clean EasyOne Air and to disinfect your hands after each patient, do the
following:
a) Use a soft cloth with a cleaning solution according to the list of cleaning
solutions under Hygiene and cleaning, 80 and wipe the handles and
the display of EasyOne Air.
b) Put down EasyOne Air.
c) Take off or change the disposable gloves and disinfect your hands
before you put on new disposable gloves.
Calibration check, 118
9.2 List of cleaning solutions for EasyOne Air
Do not use just any cleaning solution for EasyOne Air. Only use the listed
cleaning solutions.
CAUTION
ELECTRIC SHOCK, FAILURE OF EASYONE AIR, AND FALSE DIAGNOSIS
Make sure that no fluid penetrates the inside of EasyOne Air while cleaning.
To clean EasyOne Air, do not immerse EasyOne Air in any fluid.
To clean EasyOne Air from environmental dirt, you can use a damp cloth.
After each patient, clean EasyOne Air. You can use a soft cloth with a cleaning
solution according to the following list.
The cradle may be cleansed with the same cleaning solution.
70 percent isopropyl alcohol solutions or
Solutions that contain formaldehyde (only for use outside the USA) or
Solutions that contain chlorine (e.g. Sani-Cloth Germicidal Disposable
Wipes, EPA Reg# 9480-4)