EasyManua.ls Logo

NDD EasyOne Air - Page 110

NDD EasyOne Air
144 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuration
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch 110/144
117
GOLD/Hardie interpretation algorithm
The diagram above shows how the interpretation for GOLD/Hardie 2003 and
GOLD/Hardie 2008 is determined, according to GOLD (2003) [11] and Hardie
(2002) [12].
The smoker status is not part of the GOLD/Hardie standard of 2008, only of
2003.
GOLD/Hardie 2003
only
GOLD/Hardie Interpretation 2003 & 2008
START
No interpretation
No
Yes
FEV
1
/FVC limit age related:
age < 70 years: 70%
\HDUV˖DJH\HDUV
DJH!\HDUV
Maneuver
quality OK?
Obstruction
Normal spirometry
Very severe
obstruction
END
FVC < LLN and
FEV
1
< LLN
No
Yes
Yes
Yes
Yes
No
No
No
No
Yes
Possible restriction
Restriction
FEV
1
/FVC
< FEV
1
% limit?
FEV
1
˗
predicted?
FEV
1
˗
predicted?
FEV
1
˗
predicted?
Mild
obstruction
Moderate
obstruction
Severe
obstruction
Normal
spirometry—
at COPD risk
Smoker?
No
Yes

Table of Contents

Related product manuals