Calibration check
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch 119/144
121
Multi-flow cal check
With multi-flow cal check, you simulate one inspiration and one expiration for
three different flow levels each:
3 L in approximately 0.5 s (flow rate of 6 L/s)
The trial is accepted if the parameter FEF
25-75
or FIF
25-75
is between
5.50 L/s and 6.50 L/s.
3 L in approximately 3 s (flow rate of 1 L/s)
The trial is accepted if the parameter FEF
25-75
or FIF
25-75
is between
0.75 L/s and 1.25 L/s.
3 L in approximately 6 s (flow rate of 0.5 L/s)
The trial is accepted if the parameter FEF
25-75
or FIF
25-75
is between
0.40 L/s and 0.75 L/s.
To pass the complete calibration check, there must be one acceptable trial for
each flow level.
The printout shows at least three trials, one for each flow level.
Linearity cal check
The procedure for linearity cal check is the same as for multi-flow cal check,
except that three acceptable trials for each flow level are required.
The printout shows at least nine trials, three for each flow level.
List of tests and parameters, 36

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the NDD EasyOne Air and is the answer not in the manual?

NDD EasyOne Air Specifications

General IconGeneral
BrandNDD
ModelEasyOne Air
CategoryMedical Equipment
LanguageEnglish

Related product manuals