Calibration check
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch 121/144
121
7 Choose > > Single flow, Multi flow, or Linearity and confirm the
message that is displayed.
8 Wait for the baseline to be set.
9 Pull the piston all the way out as slow or as fast as required by the flow level
that is displayed.
10 Push the piston all the way in again with the same speed.
11 To end the maneuver, pull the piston briefly again.
If the maneuver was not acceptable, you need to repeat steps 9 to 11.
If you have chosen Single flow before and the maneuver was
acceptable, the calibration check is completed.
If you have chosen Multi flow or Linearity before and the maneuver was
acceptable, proceed with step 12.
12 If you have chosen Multi flow or Linearity before and the previous
maneuver was acceptable, perform the next maneuvers according to the
messages that are displayed.
If a maneuver was not acceptable, you need to repeat steps 9 to 11.
If all maneuvers were acceptable, the calibration check is completed.
List of troubleshooting solutions, 122
Contact information, 4

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the NDD EasyOne Air and is the answer not in the manual?

NDD EasyOne Air Specifications

General IconGeneral
BrandNDD
ModelEasyOne Air
CategoryMedical Equipment
LanguageEnglish

Related product manuals