Specifications and bibliography
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch 126/144
133
Instrument
classification
Type BF applied part
Powered internally with a lithium-ion battery
Short time operation, less than 10 minutes
Instrument not suitable for use in areas with flammable anesthetic gases
and in areas with mixtures containing O
2
or NO
Wireless connection This specification is not applicable to the EasyOne Air TL variant.
Type Bluetooth: v2.1 + EDR, v3.0 + HS,
v4.0
WiFi : IEEE 802.11a/b/g/j,
802.11n, 802.11d/e/i/w, 802.1X,
802.11k/r/v, 802.11h
Modulation Bluetooth: GFSK, DQPSK, 8DPSK WiFi: OFDM with BPSK, QPSK,
16-QAM, 64-QAM 802.11b with CCK
and DSSS
QoS WMM and WMM Power Save Support
Frequency band Bluetooth: 2.402 GHz to 2.480 GHz WiFi: 2412 MHz – 2484 MHz
and 4910 MHz – 5825 MHz
Transmit power Bluetooth: 15 dBm (class-1) WiFi: 18dBm
Receive sensitivity Bluetooth: -94dBm WiFi: -97dBm
Contains
transmitter module
FCC ID: XF6-RS9113SB
IC ID: 8407A-RS9113SB
Storage conditions Temperature -20°C to 50°C (-4°F to 122°F)
Relative humidity 5% to 90%
Atmospheric
pressure
500 hPa to 1060 hPa
Battery removed battery
Operating conditions Temperature 0°C to 40°C (32°F to 104°F)
Relative humidity 5% to 90%
Atmospheric
pressure
700 hPa to 1060 hPa
Specifications for EasyOne Air