Performing spirometry
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch 62/144
74
To end the test
1 To remove the EasyOne Flow Tube, do the following:
a) Grip EasyOne Air with both hands, with the backside of EasyOne Air
facing towards you and with the EasyOne Flow Tube pointing up.
b) Press against the protruding lip of the EasyOne Flow Tube with both
thumbs and push the EasyOne Flow Tube out.
c) Do not touch the EasyOne Flow Tube anywhere else than at the
protruding lip.
d) Drop the EasyOne Flow Tube directly into a waste container.
2 WARNING! Patient cross-contamination and infection of the technician:
Always wear disposable gloves. Between patients, always exchange
disposable gloves, clean EasyOne Air, and disinfect hands. / Electric
shock, failure of EasyOne Air, and false diagnosis: Make sure that no fluid
penetrates the EasyOne Flow Tube holder or the inside of EasyOne Air while
cleaning. To clean EasyOne Air, do not fully dip EasyOne Air into any fluid.
To clean EasyOne Air and to disinfect your hands after each patient, do the
following:
a) Use a soft cloth with a cleaning solution according to the list of cleaning
solutions under Hygiene and cleaning, 80 and wipe the handles and
the display of EasyOne Air.
b) Put down EasyOne Air.
c) Take off or change the disposable gloves and disinfect your hands
before you put on new disposable gloves.
3 If you want to print the report immediately, choose Print.
A progress bar is displayed that indicates the progress of the print-out.