LDS1000 Controller for Laser Diode Sensors
For all the parameters, a modification menu uses the following functions:
(F1)Scrolling of possible parameters (or cursor movement).
(F2)Lower value (-1 for digital values).
(F3)Next highest value (+1 for digital values).
(F4)Next parameter.
(F5)Save and exit mode.
List of parameters:
The values displayed on the screen of the controller are obtained by the
following calculation:
Y
Displayed
=
Y
Calculated
x(Ycoef x10 xYincr)
10
Z
Displayed
=
Z
Calculated
x(Zcoef x10 xZincr)
10
Designation Available Values Default Value
S/N Series number of 6 digits: 000000
the optical head 000000 à 999999
Type Sensor type Y axis horizontal (TR) RT
or vertical (RT)
Ycoef Linear correction 5 digits: 04000
factor for Y axis 00100 to 29999
Yincr Counter increment 0.1; 0.2; 0.5; 1; 2; 5; 1
for Y axis 10; 20; 50
Yunit Unit displayed µrad; sec; mrad; nm; unit
for Y axis µm; mm; unit
Yofs1 Electric offset 5 digits: 00000
Yofs2 values for Y 00000 to 32767
Zcoef Linear correction 5 digits: 04000
factor for Z axis 00100 à 29999
Zincr Counter increment 0.1; 0.2; 0.5; 1; 2; 5; 1
for Z axis 10; 20; 50
Zunit Unit displayed µrad; sec; mrad; nm; unit
for Z axis µm; mm; unit
Zofs1 Electric offset 5 digits: 00000
Zofs2 values for Z 00000 to 32767