49
Control panel
View
The View settings toggles to display or not the SELECT system info in the status
panel.
System info – check or uncheck the box next to “System info” to toggle ON or
OFF the display of system information within the status panel.
Figure 34 CONTROL settings
TBD
Control panel
The Control panel features three Operating modes:
• Normal – Up to 8 constant wavelengths can be configured for each crystal.
• Fast Wavelength Switching – wavelengths are switched based on settings
and signaling from an external application with a COMMAND module.
• Wavelength Sweep – the SELECT steps through multiple wavelengths at a
set rate.
Normal Operating
mode
The Control panel displays the SELECToutput channel settings for the selected
AOTF crystal. For each crystal (VIS/NIR or NIR/IR output), up to 8 wavelengths
and their individual power levels are set using the channel buttons. Select and
configure a crystal’s channels by clicking on the Crystal drop down menu, see
Figure 35.
Note: If the system is equipped with a single External RF Driver, only a single crys-
tal or output is available for operation. You can select either AOTF using the Crys-
tal selection drop down menu under Channels.
Check to toggle ON displaying System info in the status panel