Table 8 Operating and storage environment
Table 9 Safety and regulatory compliances
Figure 43 Mechanical dimensions – Select module
Specification Select Module External RF Driver
Operating Temperature
18°C to 28°C 15°C to 37°C
Storage Temperature
-20°C to 50°C N/A
Operating Humidity (non-condensing)
20 to 80% RH 20 to 80% RH
Operating Altitude
3000 m maximum N/A
Operating Air Pressure 700 hPa to 1060 hPa N/A
Safety Regulatory
EN 60825-1:2014: Safety of laser products
Part 1: Equipment classification and requirements
[Laser Class 4]
EN 61010-1:2010:Safety requirements for electrical
equipment for measurement, control, and laboratory use
Part 1: General requirements
2014/30/EU: EMC directive
2014/35/EU: Low Voltage directive
EN 61326-1:2013: Electrical equipment for measurement,
control and laboratory use
EMC requirements – Part 1: General requirements