EasyManua.ls Logo

NKT Photonics SUPERK SELECT - Table 8 Operating and Storage Environment; Table 9 Safety and Regulatory Compliances; Figure 43 Mechanical Dimensions - Select Module

NKT Photonics SUPERK SELECT
77 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
58
Table 8 Operating and storage environment
Table 9 Safety and regulatory compliances
Figure 43 Mechanical dimensions – Select module
Specification Select Module External RF Driver
Operating Temperature
18°C to 28°C 15°C to 37°C
Storage Temperature
-20°C to 50°C N/A
Operating Humidity (non-condensing)
20 to 80% RH 20 to 80% RH
Operating Altitude
3000 m maximum N/A
Operating Air Pressure 700 hPa to 1060 hPa N/A
Safety Regulatory
EN 60825-1:2014: Safety of laser products
Part 1: Equipment classification and requirements
[Laser Class 4]
EN 61010-1:2010:Safety requirements for electrical
equipment for measurement, control, and laboratory use
Part 1: General requirements
2014/30/EU: EMC directive
2014/35/EU: Low Voltage directive
EN 61326-1:2013: Electrical equipment for measurement,
control and laboratory use
EMC requirements – Part 1: General requirements
2

Table of Contents