Section 6 - Operating your Moisture Analyzer
6.5 Selecting the switch-off criterion
This menu offers you different switch-off criterion. Switch-off criterion defines when the instrument should end the
drying. Switch-off criterion eliminates checking your watch or clock and stopping the drying manually. Review the list
below and select one criterion that suits your purpose.
The following settings can be selected for the switch-off crtierion:
Manual switch-off
Timed switch-off
Auto switch-off (weight loss per unit of time - 3 settings)
Auto free switch-off (user-defined mean weight loss per unit of time)
TEST PARAMETERS
Manual
With this switch-off criterion selected, the measurement process continues
until you stop it with the Stop button.
Using the arrow buttons, scroll to SWITCH OFF and press the Enter button.
SWITCH OFF
Using the arrow buttons, scroll to MANL - USE START/STOP and press the
Enter button.
SWITCH OFF
Timed switch-off
With this switch-off criterion selected, the measurement lasts until the preset
drying time has elapsed (the display provides you with continuous informa-
tion on the drying time).
Using the arrow buttons, scroll to SWITCH OFF and press the Enter button.
Using the arrow buttons, scroll to TIMED - SWITCH OFF and press the Enter
button.
TIMED SWITCH OFF
Using the arrow buttons, enter the desired drying time and press the
Enter button.
EXIT WITHOUT SAVING
EXIT WITHOUT SAVING
EXIT WITHOUT SAVING
TEST ID: DEFAULT
DRY TEMP: 125C
PROFILE: STANDARD
SWITCH OFF: MANUAL
RESULT: MOISTURE%
CUSTOM : OFF
MANL - USE START/STOP
TEST PARAMETERS
TEST ID: DEFAULT
DRY TEMP: 125C
PROFILE: STANDARD
SWITCH OFF: TIMED 20.00
RESULT: MOISTURE%
CUSTOM : OFF
TIMED - SWITCH OFF
20:OOMIN