59
Section 9 - Printing
9. Printing test data
Printing data to an external computer or printer requires the communication parameters be set first.
Printing to an external printer or computer will occur each time the Print button is pressed. If the Print Interval is set in
the Test Menu, printing can occur in a continuous fashion at specified intervals.
Samples of weight calibration, temperature calibration and test data (GLP on and GLP off) are shown as follows:
WEIGHT CALIBRATION #12
DATE: 12.19.01
TIME: 11:20
OHAUS MB45
S/N: 1234567
REFERENCE WT: ...................
WEIGHT: 20.000 G
CALIBRATION PERFORMED
SIGNATURE:
[..........................................]
[------------END------------------]
Weight calibration printout
TEMPERATURE CAL#12
DATE: 12.19.01
TIME: 11:25
OHAUS MB45
S/N: 1234567
TEMP REFERENCE ID..........
TEMP 100C: 103C
TEMP 160C 161C
CALIBRATION PERFORMED
SIGNATURE:
[...........................................]
[------------END-------------------]
Temperature calibration printout