67
Section 10 - Care and maintenance
MODEL MB45
Capacity 45g
Readability 0.001g, .01%
Results Moisture (%), dry content (%), Wt (g), Wt (custom),
regain, %
Temp. Settings 50°C to 200°C
Increments 1°C
Drying Programs Standard, fast, ramp, step
Switch-off Criteria Timed, Auto (30, 60, 90 seconds, auto free), manual
Heat Source Halogen
Display (in/cm) 2.75 x 1.50 / 7 x 3.8
Display Type LCD- 128 x 64 Pixels
Calibration External calibration mass-20g
Dimensions (DxWxH) (in/cm) 14 x 7.5 x 6/ 35.5 x 19 x 15.2
Pan Size 90 mm. diameter
Adjustable Feet and Level Yes
RS232 Interface Yes
Weight (lb/kg) 9.8 / 4.5
Shipping Weight (lb/kg) 14.1 / 6.4
10.7 Specifications
Admissible ambient conditions
Use only in closed rooms
Temperature range: 5 ºC to 40 ºC
Atmospheric humidity: 80% rh @ to 30 ºC
Warm-up time: At least 60 minutes after connecting the instrument to the power supply;
when switched on from standby-mode, the instrument is ready for
operation immediately.
Voltage fluctuations: 15% +10%
Installation category: II
Pollution degree: 2
Power load: Max. 450 W during drying process
Current consumption: 4 A or 2 A, according to the heating element
Power supply voltage: 100 V 120 V or 200 V 240 V, 50/60 Hz
(the voltage is given by the heating element)
Power line fuse: 1 piece, 5 x 20 mm, T6,3 H 250 V