116
Chapter 5 Functions
EVIS EXERA II VIDEO SYSTEM CENTER CV-180
5. Click “Zoom”. The message “Please wait” is displayed on the monitor, then
the full screen image appears (see Figure 5.66).
Figure 5.66
6. Press the “Esc” key to return to the normal sized playback screen.
Or press the “Shift” and “F4” keys together to return to the endoscopic
image screen.
• The images recorded by an instrument other than the CV-180
cannot be played back.
• The images edited by a personal computer or an instrument
other than the CV-180 may be unable to be played back.
• The endoscopic image and the patient data is displayed
together on the same screen when using the CV-180 for
playing back, but cannot be displayed together on a personal
computer.
• In the normal screen, all images recorded in the full-height or
full size are scaled down and displayed on the constant
display area are shown in Figure 5.65.
• The two images of the PinP are displayed separately in the
thumbnail screen, which has the same time stamps but
different file names.
• The sub image of the PinP is displayed with the same patient
data as the main image.
• Playback in PinP mode is not available.
• “R” is not displayed on the monitor when the rotated image of
the orientation function is played back (refer to “Monitor
orientation function” on page 245).