188
Chapter 8 Installation and Connection
EVIS EXERA II VIDEO SYSTEM CENTER CV-180
EU-M60, EU-MA
For connecting the ultrasound center (EU-M60 or EU-MA) to the video system
center, use the cable listed below.
Figure 8.26
Do not connect the video output and B/W output of the EU-
M60 to the CV-180. An ultrasonic image in the PinP display
may be distorted.
Model Product name Note
MH-985 or MAJ-987 S cable -
Table 8.24
CV-180
EU-M60, EU-MA
1
PinP Y/C
VTR OUT
MH-985
or
MAJ-987
1 The terminal layout of EU-MA is
differ slightly from that of this
illustration.