EasyManua.ls Logo

Olympus CV-180 - Page 206

Olympus CV-180
296 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
198
Chapter 9 Function setup
EVIS EXERA II VIDEO SYSTEM CENTER CV-180
Light sources
This operation sets the type of light source connected.
Other settings of the light source may cause improper color
and/or brightness in the image.
1. Click “ ” of the list box “Type”. The following light sources appear in the
pull-down menu.
2. Click the setting value corresponding to the connected light source. The
selected setting value is displayed.
• When the CLV-180 is used, the setting is not necessary.
• When a light source other than CLV-180 is used, set up the
light source according to the table below. Please refer to the
instruction manual of the light source.
Setting value Connected light source
CLV CLV-160, CLV-U40
CLV-S CLV-S40
CLE Halogen light source
Table 9.4
Setting items Settings
Brightness adjustment mode AUTO
BRIGHTNESS Center of the adjustable range
Table 9.5

Table of Contents