EasyManua.ls Logo

Olympus CV-180 - Page 275

Olympus CV-180
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 10 Troubleshooting
267
EVIS EXERA II VIDEO SYSTEM CENTER CV-180
When the following messages appear, contact Olympus.
VCR error 1 The VCR remote cable is
not connected correctly.
Connect the VCR remote
cable correctly.
VCR error 2 Incompatible format of
media
Use a compatible media
referring to the instruction
manual for the media.
Message Possible cause
PC Card Error 01 Failure to call on the image from the PC card
PC Card Error 02 The number of releases for a folder exceeds 9999
PC Card Error 03 Failure to delete the image
PC Card Error 04 Failure to call on the image management file
PC Card Error 05 Failure to write to the image management file
PC Card Error 06 Failure to call on the folder
PC Card Error 07 The number of releases for the PC card exceeds 900
PC Card Error 08 Failure to delete the folder
PC Card Error 09 Failure to call on the annotation file
PC Card Error 10 Failure to create the annotation file
PC Card Error 11 Failure to delete the annotation file
PC Card Error 12 Failure to call on the image attached to the annotation file
PC Card Error 13 Failure to write the image to the annotation file
PC Card Error 14 Failure to call on the patient data
PC Card Error 15 Failure to create the patient data
PC Card Error 16 Failure to call on the setting file
PC Card Error 17 Failure to create the setting file
PC Card Error 18 Failure to create the log file
PC Card Error 19 Failure to compress the file
PC Card Error 20 Failure to decompress the file
PC Card Error 21 Released with full media
Message Possible cause Solution

Table of Contents