EVIS EXERA II VIDEO SYSTEM CENTER CV-180
Cautions and recommended electromagnetic environment
regarding portable and mobile RF communications
equipment such as a cellular phones
• Where “P” is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and “d” is the recommended separation
distance in meters (m).
• This instrument complies with the requirements of
IEC 60601-1-2: 2001. However, under electromagnetic
environment that exceeds its noise level, electromagnetic
interference may occur on this instrument.
• Electromagnetic interference may occur on this instrument
near a high-frequency electrosurgical equipment and/or other
equipment marked with the following symbol:
Immunity test
IEC 60601-1-2
test level
Compliance
level
Guidance
Conducted RF
IEC 61000-4-6
3V
rms
(150 kHz - 80 MHz)
3V (V
1
) Formula for recommended separation distance
(V
1
=3 according to the compliance level)
Radiated RF
IEC 61000-4-3
3V/m
(80 MHz - 2.5 GHz)
3V/m (E
1
) Formula for recommended separation distance
(E
1
=3 according to the compliance level)
80 MHz to 800 MHz
800 MHz to 2.5 GHz