EasyManua.ls Logo

Olympus CV-180 - Page 50

Olympus CV-180
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42
Chapter 4 Operation
EVIS EXERA II VIDEO SYSTEM CENTER CV-180
If any other abnormality occurs or is suspected,
immediately stop using the equipment, turn OFF all
equipment, and gently withdraw the endoscope from the
patient as described in the endoscope's instruction
manual. Then refer to the instructions in Chapter 10,
“Troubleshooting”. If the problems cannot be resolved by
the remedial action described in Chapter 10, do not use
the equipment and contact Olympus.
Combination with other equipment
Do not use the video system center in locations exposed
to direct strong electromagnetic radiations (for example,
microwave treatment device, short wave treatment
device, MRI or radio equipment). Electromagnetic
radiation can interfere with the monitor display.
Use only Olympus high-frequency electrosurgical
equipment with this unit. Non-Olympus equipment can
cause interference on the monitor display or a loss of the
endoscopic image.
Before using high-frequency electrosurgical equipment,
be sure to install and connect the equipment according to
it’s instruction manual and make sure that the noise does
not affect the observation and surgical procedures. If
high-frequency electrosurgical equipment is used without
such confirmation, patient injury may result.
To activate the auto brightness control function of the light
source, the video system center should be turned ON. If it is
not turned ON, the auto brightness control function is not
activated and the light intensity is set to maximum. In this
case, the endoscope distal end would become hot and could
cause burns to the operator and physician (if a light source
model other than CLV-180 is used).
When using spray-type medical agents such as lubricant,
anesthetic, or alcohol, use them away from the video system
center so that the medical agents do not contact the video
system center. Medical agents might enter the video system
center through the ventilation grills and may cause
equipment damage.
Do not use a humidifier near the video system center as dew
condensation possibly might occur and it may cause
equipment failure.

Table of Contents

Related product manuals