5 Setup and Wiring
5 - 26
Vision System FH/FZ5 series Hardware Setup Manual (Z366)
Precautions for Safe Use
Power Supply and Wiring
Keep the power supply wires as short as possible (Max.10 m).
Use the cables and crimping terminals with the sp
ecified dimensions. Do not directly connect
an electric wire to the terminal lock that is simply twisted.
Recommended wire size: 1.31 to 2.63 mm
2
Terminal screw: M4 (Tightening torque:1.4 N·m)
Crimping Terminal
Ground
The controller power circuit is not insulated from its internal circuit.
When the connected camera to the Sensor Control comes packaged with a base, make sure
to mount w
ith the base. As the housing of the camera is the SG (0 V), it can cause short-cir-
cuiting between the SG (0 V) and the FG if a spacer is not used.
It is short-circuited with FG of the customer dev
ice when installing it directly because the
case of the controller is connected with SG (0 V).
Do not ground the 24 VDC power supply's posit
ive terminal. If the positive terminal is
grounded, it causes the electric shock when you touch the SG (0 V) such as case of the con-
troller or the camera.
Use the cables and crimping terminals with the specified dimensions.
Keep the power supply wires as short as possible (Max.10 m).
Wire the power supply unit inde
pendently of other devices.
Do not directly connect an electric wire to t
he terminal block that is simply twisted.
After wiring, replace the terminal cover.
Power source types for FZ5-L series differ depending on the number of cameras due to current con-
sumption differences. Refer to the following table to use the ap
propriate type. When you connect your
camera to the lighting via Light Controller, the current consumption is same as when the Intelligent
Compact Digital camera is connected.
FZ5-L Series
5-3-5 FZ5-L Series
Connection of Terminal Block of FZ5-L Series
Recommended Power Source for FZ5-L Series
Item Camera Type
Number of
Camera
FZ5-L35 FZ5-L35-10
Recommended Source
S8VS
Intelligent Compact
Di
gital Camera
2 S8VS-12024 S8VS-18024
4 ---
Camera of 0.3/2/5
mill
ion pixels
2 S8VS-09024 S8VS-09024
4 ---
8.5 mm max. 8.5 mm max.

Table of Contents

Other manuals for Omron SYSMAC FH Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron SYSMAC FH Series and is the answer not in the manual?

Omron SYSMAC FH Series Specifications

General IconGeneral
BrandOmron
ModelSYSMAC FH Series
CategoryIndustrial Equipment
LanguageEnglish

Related product manuals