Operator’s Manual – OPTI CCA-TS2 9-15
9 OPERATING PRINCIPLES
CalibrationGas,BP7162
Use: For calibration of pH, PCO
2
and PO
2
in the OPTI CCA-TS2 Analyzer.
Contents: Each disposable, low-pressure cylinder contains 0.35 liters of gas
at 28 psi at 21 °C.
Composition: Oxygen 14.0 ± 0.02%
Carbon Dioxide 6.0 ± 0.02%
Nitrogen balance
Storage: Refer to package labeling.
HbCalibratorCassette,BP7653
Use: For quarterly calibration of the OPTI CCA-TS2 Analyzer.
Contents: Each package contains one reusable calibrator cassette.
Composition: Stabilized optode sensors
Storage: Refer to package labeling.
Stability: Expiration date and lot number are printed on each package label and
encoded on the attached bar code label.
Precautions
Use of calibration solutions, calibration gas, or optodes not manufactured by OPTI Medical
Systems could void the warranty.
Once used, the sample cassette holds human body uids which may be potentially infectious;
handle with appropriate care to avoid skin contact or ingestion.
For in-vitro diagnostic use.
For professional use only.
Bibliography
1. Guilbault GG, Ed., PracticalFluorescence, 2nd Edition, Marcel Dekker, 1990
2. Tietz; Burtis C, et al (Eds.), TextbookofClinicalChemistryandMolecularDiagnostics, 4
th
Ed.,
(Elsevier Saunders, 2006) pps. 2252-2302.
3. Mahoney JJ, Wong RJ, Van Kessel AL, ReducedBovineHemoglobinSolutionEvaluatedforUseasa
BloodGasQualityControlMaterial. Clin.Chem.39/5,874-79 (1993)
4. J.B.Henry, ClinicalDiagnosisandManagementbyLaboratoryMethods, 19th Ed., 1996, p.590,777
5. Clinical and Laboratory Standards Institute (CLSI). InterferenceTestinginClinicalChemistry;
ApprovedGuideline-2ndEdition. CLSI document EP7-A2. CLSI, Wayne, PA, 2005